Measuring Range; Cell Constant: |
Y |
Measured .01006 1/cm, Nominal .01 1/cm | ||
Process Connection: |
9Y |
G1 or 1" | ||
Cable connection: |
5 |
- |
- |
|
Temperature Sensor: |
A |
- |
- |
-
Measuring principle
Conductive
-
Application
Power plant, condensate, boiler feed water
-
Characteristic
2-electrode conductivity cell for high temperature applications in pure and ultrapure water
-
Measuring range
k=0.01: 0 to 20 µS/cm
k= 0.1: 0 to 200 µS/cm -
Measuring principle
Conductive conductivity cell with polished stainless steel electrodes
-
Design
Large coaxially arranged electrodes for a fast media flow through the cell
-
Material
Electodes: stainless steel 1.4571
Sealing: PEEK / EPDM -
Dimension
Electrode diameter: 25 mm (0.98 inch)
Immersion depth: 50 to 115 mm (1.97 to 4.53 inch) -
Process temperature
-30 to 160 °C (-22 to 320 °F)
-
Process pressure
Max. 40 bar up to 100 °C (max. 580 psi up to 212 °F) without CLA 751 flow chamber
Max. 12 bar up to 150 °C (max. 174 psi up to 302 °F) with CLA 751 flow chamber -
Temperature sensor
Integrated Pt100
-
Ex certification
ATEX
-
Connection
Process connection : G1" and NPT1" Fixed cable on transmitter
-
Ingress protection
IP67
-
Additional certifications
Calibration certification with cellconstant