Material; Solution Ground: |
30 |
|||
Flow In; Out: |
A |
|||
Process Connection: |
C |
|||
Seal: |
2 |
|||
Additional Option: |
10 |
-
Measuring principle
Potentiometric
-
Application
Drinking water treatment plants, boiler feedwater and ultra-pure water plants, cooling-water circuits, fertiliser production, sugar production, gas scrubbers, petrochemical plants
-
Installation
Flow assembly
-
Characteristic
Piping, bypass
-
Design
3 electrode installation locations, 120mm electrodes, flow direction horizontal or from the bottom, thread connection G1/2" for spray cleaning
-
Material
Flow vessel: PVDF or stainless steel 1.4404
O-rings: EPDM, Viton, Chemraz or Fluoraz -
Process temperature
PVDF version: 0 to 120 °C (32 to 248 °F)
Stainless steel version: -15 to 150 °C (5 to 302 °F)
With EPDM: -15 to 140 °C (5 to 284 °F) -
Process pressure
PVDF version: max. 8 bar at 50 °C (max. 116 psi at 122 °F)
Stainless steel version: max. 10 bar (max. 145 psi) -
Connection
Pressurized flange DN25/PN16, ASME 1" lbs 150, JIS 10K25A
-
Additional certifications